The Pictus 700 is a fully automated, random access, small footprint floor clinical chemistry analyzer. The system can achieve a total throughput of up to 720 tests per hour.
Key Features And Benefits
Efficiency-
The Pictus 700 incorporates a 95 position patient sample tray reducing the number of required daily runs. It also has programmable auto dilution for out of range tests. The reagents are very stable requiring few calibrations thus saving time and money. The 72 reagent positions (75 w/ISE) and onboard refrigeration allow for many tests to be kept on the analyzer at the same time.
Cost Savings –
The 700 does not need an expensive water treatment that requires expensive upfront cost and continual purchase of filters. More savings will be realized by utilizing low cost reagent, calibrator and controls reducing the overall cost per reportable result. Furthermore, the extended service pricing is comparatively low versus other models of this type.
Specifications
Up to 600 tests/hour, max. 720 tests/hour with optional ISE unit
Sample volume: 2 to 100 μl/tests (in increments of 0.2 μl)
Sample tray: 95 (5 racks x 19 positions) ID barcode equipped positions for routine
Stat and control samples and standard solutions
Maximum number of simultaneous tests: 36 double to 72 single reagent tests, + 3 with optional ISE unit
1 to 3 reagents, 2 to 480 μl/tests each (in increments of 1 μl) final total solution volume 180 to 650 μl/test
Reagent bottle capacities: 72 cooled positions
Barcode Reader for reagents
Water consumption: Less than 5 l/h
Dual beam system
Automatic pre-dilution and post dilution (ratio 1:5 to 1:100)
Full quality control: Levey-Jennings plots, Westgaard multi-rules
Windows-based software
Power requirements: 110/230 V, 50/60 Hz, 1.4 kVA
Size (Width x Depth x Height): 97 cm x 67 cm x 100 cm, 38” x 26” x 39”
Weight: 190 kg, 419 lbs
Options
ISE Unit - Na+, K+ and CI- measurements. Other electrolytes on request.
PC Requirements- PC compatible with Celeron 2.4 GHz, 256 MB RAM, 40GB Hard Disc or bigger, CD ROM, Super VGA 15” Color screen, with Microsoft Windows™ 8.1 or 10 Operating system.
Printers - Most printers supported by Windows can be connected. Printouts can be optimized by the customer. Results analysis, work list, patient sample list, quality control, calibration curves and more.